Fractured Leader Italia implants: thoughts?

Last Updated: Apr 21, 2014

I am a Spanish Dentist. I placed Leader Italia, Tixos internal implants, a few years ago. Three implants on two of the patients have now fractured. In both cases, the patients had strong bite forces, although the protheses were different (a superior hybrid with two implants fractured and a second maxillar premolar with two crowns -without occlusion over the extension). Has anybody had a similar experience with the Leader Italia implants? What are your thoughts on this?

5 Comments on Fractured Leader Italia implants: thoughts?

Vasileios Kouvas

04/22/2014

Dear colleague from Spain I have the same bad experience with two TIxos internal hexagon implants from Leader Italia.The implant position was in 21,22 and 26 with a cement -retained metal -porcelain bridge.Three years later,my patient came to my office with complaints because of mobility of the restoration.The -x ray examination revealed that the broken implants were in the position of 22 and 26.The fracture line was at the same level of the apical level of the prosthetic screw.There was no evidence of strong chewing habits or bruxism.As you can supposed , the cost of the new implants and the prosthetic restoration was mine.. Until now I have the fear of the recurrence because I have finished a lot of cases with Tixos implants from Leader

Dennis Flanagan DDS MSc

04/29/2014

I have completed an in vitro research project on this issue. We found that implants can fracture at the cervical when loaded with more than 250 N in an off axial direction. This was also found in proprietary research by 3-I, poster presented at the AO meeting. It is a good idea to measure a pts bite force with pressure indicators from TekScan in South Boston, Mass. USA. This information is helpful in treatment planning.

Jaime

04/30/2014

1) Were machined components used ? 2) Don´t forget , even when the metal try-in in 2-dimensions looks great on radiograph , after firing porcelain , we have distortion . Southern Implants have passive abutments . 3) In case of high bite force keep the occlusal table narrower and decrease the inclined planes of the cusps .

José Vilches

05/05/2014

Hi. Abutments were machined from Leader. The fracture was in the middle of the implant, no in the platform, so I think would be a problem in the manufacurer.

José Vilches

10/29/2014

Hi. Another Tixos fractured. A #22 with a bridge screwed.

Featured Products

DALI Bone Mix

DALI Bone Mix

The highest quality tissue!

Classic

Classic 50/50 Mix

Promotes osteoconduction

Provides structural integrity

DALI Bone Syringe

DALI Bone Syringe

Prefilled Mineralized Cortico-Cancellous Bone in Syringe

New

Convenient Syringe!

50/50 Cortical/Cancellous

Available in 3 sizes.

Osteogen Plug

Osteogen Plug

Combines bone graft with a collagen plug.

Classic

Eliminate hassle of mixing particulate grafts

Sold in packs of 5 or packs of 10.

Proven safe, and clinically effective

OsseoSeal Flexible Membrane

OsseoSeal Flexible Membrane

Resorbable collagen membrane derived from purified porcine pericardium

Popular

Fast hydration and excellent tensile strength

Good adaptation to various defects

Excellent tear function and duration

DALI One Graft

DALI One Graft

One-Step grafting solution!

New

100% allograft

Eliminates mixing hassle

Moldable after hydration